The University of Texas Health Science Center - Graduate School of Biomedical Science
University: Home | Calendar | Maps
We Make Lives Better

Site University
  • Home
  • About Us
    • Eileen M. Lafer, Ph.D.
    • Virgil Schirf, MS
  • Publications
  • Fee Schedule
  • Forms
    • Biacore Experiment Checklist (pdf)
    • Client Acknowledgment Form (pdf)
  • Please read before you start
    • Selected Literature
    • Training
    • Before we start
    • Rmax Calculator
    • Sensor chips
    • Chemical compatibility
    • Injection information
    • Wash Commands
    • Maintenance
    • BIA programs
  • Protocols
    • HBS-EP running buffer
    • Pre-concentration (pH scouting)
    • Pre-conditioning of sensor chip
    • Activation, ligation, and blocking
    • Minimal Biotinylation
  • Service
    • Service/PM September 17-18, 2009
  • UTHSCSA
    • UTHSCSA
    • Cancer Therapy and Research Center
    • Department of Biochemistry
    • Research @ UTHSCSA

News

  • Health Science Center Headlines

Calendars

  • Biochemistry Seminar Schedule
  • UTHSCSA Calendar of Events
Skip Navigation

UTHSCSA Center for Surface Plasmon Resonance (CSPR)

Types of Injections and Volumes

General purpose inject
Command Comments Injected volume Consumption of
sample
Max flow rate Inj Volume Prep Time Finish Time Air
Segments
INJECT 5 - 325 µl Injected vol + 30 µl 100 µl/min 20
100
80
90
50
50
2
KINJECT Low dispersion inject for kinetics and for samples sensitive to dilution with running buffer. Controlled dissociation time after sample injection. 10 - 250 µl Injected vol + 40 µl 100 µl/min 20
100
40
45
20
25
1
QUICKINJECT Low sample consumption. Less preparation time than INJECT but more sample dispersion. 5 - 325 µl Injected vol + 10 µl 100 µl/min 20
100
125
125
40
45
3
CONINJECT Injection of second sample immediately following the first e.g. for dissociation in the presence of ligand. Variable contact times for the first sample only. 10 - 100 µl for first sample
15 - 250 µl for second sample
Injected vol + 40 µl for first and second sample 100 µl/min 20,20
100,100
200
250
80
80
3
BIGINJECT Large volume inject. Direct input up to 325 µl, sample loop between 326 and 425 µl, direct between 426 and 750 µl. 426 - 750 µl Injected vol + 52 µl 50 µl/min 350
700
400
400
90
100
3
MANUAL INJECT Manually controlled injection 1 - 100 µl loading Loaded volume + 30 µl 100 µl/min Prep time for loop
loading 100 µl:80s
85 2
 
Department of Biochemistry | Graduate School of Biomedical Sciences | UTHSCSA

UTHSCSA | Graduate School of Biomedical Sciences | Department of Biochemistry | CTRC | Research

Links from web sites affiliated with The University of Texas Health Science Center at San Antonio's web site (www.uthscsa.edu) to other web sites do not constitute or imply university endorsement of those sites, their content, or products and services associated with those sites.

Contact: Webmaster
Updated: 10/15/2009
Copyright © 2007
The University of Texas Health Science Center at San Antonio