The University of Texas Health Science Center - Graduate School of Biomedical Science
University: Home | Calendar | Maps
We Make Lives Better

Site University
  • Home
  • About Us
    • Eileen M. Lafer, Ph.D.
    • Virgil Schirf, MS
  • Publications
  • Fee Schedule
  • Forms
    • Biacore Experiment Checklist (pdf)
    • Client Acknowledgment Form (pdf)
  • Please read before you start
    • Selected Literature
    • Training
    • Before we start
    • Rmax Calculator
    • Sensor chips
    • Chemical compatibility
    • Injection information
    • Wash Commands
    • Maintenance
    • BIA programs
  • Protocols
    • HBS-EP running buffer
    • Pre-concentration (pH scouting)
    • Pre-conditioning of sensor chip
    • Activation, ligation, and blocking
    • Minimal Biotinylation
  • Service
    • Service/PM September 17-18, 2009
  • UTHSCSA
    • UTHSCSA
    • Cancer Therapy and Research Center
    • Department of Biochemistry
    • Research @ UTHSCSA

News

  • Health Science Center Headlines

Calendars

  • Biochemistry Seminar Schedule
  • UTHSCSA Calendar of Events

Core Links

  • Mass Spectrometry
  • Nuclear Magnetic Resonance
Skip Navigation

Protocol: Activation, Ligation, and Blocking

This protocol is used to activate a pre-conditioned CM sensor chip for amine coupling. The sensor chip can be used as is for amine coupling of ligands or strepavidin or neutravidin can be coupled to make a SA (NA) chip for capture of biotinylated analytes.

You will need:

  • a fresh CM chip which has been pre-conditioned
  • N-hydroxysuccinimide (NHS) which has been diluted and aliquoted. Individual 9 mm glass tubes are found in the freezer
  • 1-ethyl-3-(3-dimethylaminopropyl) carbodiimide hydrochloride (EDC) which has been diluted and aliquoted. Individual 9 mm glass tubes are found in the freezer.
  • An empty 9 mm glass tube
  • Ligand or strepavidin/neutravidin. DO NOT PREPARE IN ACETATE BUFFER UNTIL IMMEDIATELY BEFORE USE!

Procedure:

  • Dock the new chip and prime 2X with running buffer.
  • Because of the time sensitive nature of activation and capture, flow cells are done individually . Generally Flow Cell 1 (FC1) is activated and blocked immediately and used as a control surface (see figure above). Run:Start Sensorgram and choose the appropriate flow cell. The rate is set to 5 µl/min. The NHC and EDS are aliquoted and frozen. Immediately prior to use, remove one tube of each and defrost and gently mix. Place in R2E1 and R2E2 and place an empty tube in R2E3. The NHC and EDS are diluted 50:50 by the instrument and the mix is placed in the empty tube. Go to Command:Dilute and enter the information from the picture below and then click "OK".
    In the Command Queue enter Inject:Quickinject and enter a 35µl volume which allows 7 minutes of activation.
  • If you are going to use this flow cell surface as a reference, pipet 1 ml of 1 M ethanoloamine into a 16 mm glass vial, cap, and place in R2F3. Select Command:Inject:Quickinject and set the volume to 35 µl. Click on More and select Extraclean; begin injection.
  • For surfaces on which ligand is to be coupled, activate as described. Close the command queue. Select Command:Inject:Manual Inject. Load the loop. Manually inject small volumes such as 2 µl to see what response you get. Use the flags to mark on the sensorgram the baseline and to note injections. Once you have achieved teh desired RU, close the manual inject and empty the loop. Block the surface following the procedure above by injecting 1 M ethanolamine over the surface with the extracleanup step selected. Set the baseline after activation and determine the total RU of the surface after blocking. The surface below has approximately 2000 RU of protein bound to the surface after blocking.
 
Department of Biochemistry | Graduate School of Biomedical Sciences | UTHSCSA

UTHSCSA | Graduate School of Biomedical Sciences | Department of Biochemistry | CTRC | Research

Links from web sites affiliated with The University of Texas Health Science Center at San Antonio's web site (www.uthscsa.edu) to other web sites do not constitute or imply university endorsement of those sites, their content, or products and services associated with those sites.

Contact: Webmaster
Updated: 10/15/2009
Copyright © 2007
The University of Texas Health Science Center at San Antonio