
UTHSCSA Center for Surface Plasmon Resonance (CSPR)
Sample BIA method file
This is an example of a BIA method file. For your particular method you may need more or less positions for samples or may want to group all the blanks in one row. In this particular example all samples are treated the same way, with a kinject of 150 µL (1800 seconds or 30 minutes of contact time) followed by a 1800 second dissociation time and then a 5 second regeneration step from position r2f3 with extracleanup. The analytes are flowed over all 4 flow cells of the sensor chip at a flow rate of 5 µL per minute. The program loops through the steps s1..s2.........s50. Once the program ends, it tells the instrument to enter standby mode. Each cycle lasts approximately 1 hour and 8 minutes in this example. A text version is available for download and appropriate modifications. DEFINE LOOP cycle1LPARAM %s1 %p1
times 1
s1 r2a1 ! indicates sample 1 located in rack 2 (located on your right) position a1
s2 r2a2 ! generally samples r2a1 - r2a10 are buffer blanks which allows the instrument to "settle down."
s3 r2a3 ! These buffer blanks are hidden in Scrubber.
s4 r2a4
s5 r2a5
s6 r2a6
s7 r2a7
s8 r2a8
s9 r2a9
s10 r2a10
s11 r2b1 ! Lowest concentration of first analyte to flow over the sensor chip
s12 r2b2 ! Next highest concentration of the first analyte and so forth with a total of seven (7) different concentrations.
s13 r2b3
s14 r2b4
s15 r2b5
s16 r2b6 ! Highest concentration of first analyte.
s17 r2b7
s18 r2b8 ! Buffer blank to be used in Scrubbers double referencing system
s19 r2b9 ! Buffer blank to be used in Scrubbers double referencing system
s20 r2b10 ! Buffer blank to be used in Scrubbers double referencing system
s21 r2c1 ! Lowest concentration of replicate of first analyte or lowest concentration of second analyte. Repeat as above.
s22 r2c2
s23 r2c3
s24 r2c4
s25 r2c5
s26 r2c6
s27 r2c7
s28 r2c8 ! Buffer blank to be used in Scrubbers double referencing system
s29 r2c9 ! Buffer blank to be used in Scrubbers double referencing system
s30 r2c10 ! Buffer blank to be used in Scrubbers double referencing system
s31 r2d1 ! Lowest concentration of replicate of first analyte or lowest concentration of third analyte. Repeat as above.
s32 r2d2
s33 r2d3
s34 r2d4
s35 r2d5
s36 r2d6
s37 r2d7
s38 r2d8 ! Buffer blank to be used in Scrubbers double referencing system
s39 r2d9 ! Buffer blank to be used in Scrubbers double referencing system
s40 r2d10 ! Buffer blank to be used in Scrubbers double referencing system
s41 r1a1 ! indicates sample 1 located in rack 1 (located on your left) position a1
s42 r1a2
s43 r1a3
s44 r1a4
s45 r1a5
s46 r1a6
s47 r1a7
s48 r1a8
s49 r1a9
s50 r1a10
END
DEFINE APROG analyse1
PARAM %s1 %p1
flow 5 -f ! flow rate is set to 5 µL/min
flowpath 1,2,3,4 ! the analyte flows over all 4 flow cells
kinject %p1 150 1800 : ! the 150 refers to the µL of analyte to flow over the sensor chip; the 1800 refers to the number of seconds to allow for dissociation; both "on" and "off" times are 30 minutes
quickinject r2f3 5 ! regeneration with extraclean
extraclean
wash i ! wash all IFC channels
wash f ! wash flow cell area
end
MAIN
DETECTION 1,2,3,4
LOOP cycle1 STEP
APROG analyse1 %s1 %p1
ENDLOOP
append standby ! standby procedure initiated after completion of the program
END
